2 Setup methods according to DIN EN ISO 17664 | 2.7 Manual reprocessing of the water-conducting system

2.7 Manual reprocessing of the water-conducting system

Note
Using the "Water block, compact", a water bottle or the Water block DVGW assem‐
bly kit with integrated germ reducing facility but no intensive germ reduction assem‐
bly kit, the water-conducting system must be rinsed manually. The automatic hy‐
giene functions are available through the intensive germ reduction assembly kit on‐
ly.
Risk of infection if the "Water block, compact" is used without additional safeguards.
Contamination of the treatment water and/or drinking water supply with germs.
▶ With regard to the "Water block, compact" assembly kit, please note that no disin‐
fection facility is installed in the unit, and take appropriate safeguards. KaVo rec‐
ommends to use the "Water block DVGW with integrated water disinfection facility
in combination with KaVo OXYGENAL 6 (Mat. no. 04893451).
▶ If the Water bottle kit is used with the enclosed dosing attachment (Mat. no.
10020287), add the proper amount of KaVo OXYGENAL 6 (Mat. no. 04893451)
with each filling. For the correct amount, please refer to the Instructions of the
dosing attachment for water disinfection.
Manual reprocessing procedure:
▪ Before first start-up and after the device has not been used for a while (weekends,
holidays, vacation times, etc.)
▪ Before you start working
▪ After each patient (short rinsing programme of 20 seconds)
▶ Rinse all water outlets without handpieces for 2 minutes and actuate tumbler filler
repeatedly.
▶ Before first start-up and after the device has not been used for a while, briefly
blow air through the air lines.

2.8 Automatic hygiene functions of the water conducting system

Note
Intensive germ reduction kit and water block DVGW required.
Health damage due to germ formation.
Infection hazard.
▶ Before starting, rinse all the water drain lines without instruments.
▶ Before start-up and after the device has not been used for a while (weekends,
holidays, vacations, etc.), rinse or purge with air the air and water lines.
▶ Carry out an intensive germ reduction.
▶ Actuate the tumbler filler repeatedly.
The water-conducting system includes all water ducts of the treatment unit to which
the patient may be exposed. This includes the handpieces on the dentist element and
assistant element as well as the tumbler filler on the spittoon bowl of the treatment
centre.
WARNING
CAUTION
23 / 40
Care instructions Primus 1058 Life