Short instructions for use 10 Hygiene
▶ Perform the preparations for intensive germ reduction.
See also:
2 10.2 Rinsing programme, Page 27
▶ Press the "Intensive germ reduction" button.
ð Intensive germ reduction starts and runs though several steps. The process can
be followed on the display.
Intensive germ reduction is completed when "01 intensive germ reduction completed"
appears on the display.
▶ Put all handpieces in holders and take off the germ reduction attachment.
Interrupt intensive germ reduction during the dwell time
So that no one will have to remain with the device throughout the entire intensive germ reduction proc‐
ess, intensive germ reduction can be interrupted in program step 4 (e.g. on the last workday before
the weekend). Program step 4 is the dwell time, in which the microorganisms are killed by an in‐
creased OXYGENAL 6 concentration.
Requirement
intensive germ reduction is in program step 4.
You hear a beep for 4 seconds.
▶ Turn the device off.
When the device is switched on after 30 minutes of dwell time, the remaining program steps run imme‐
diately and automatically.
If the device is turned on before the 30 minute period is elapsed, the dwell times finishes first before
the remaining steps of the intensive germ reduction program run.
Skip exposure phase
This function is required when the treatment unit needs to be used (for example in an emergency, at
hospitals, etc.) even though the germ reduction procedure has already started.
Intensive germ reduction is in step 4:
▶ Press and hold down the "Hydroclean" and "Intensive germ reduction" keys.
▶ Simultaneously press the foot pedal on the foot control.
ð The intensive germ reduction switches to step 3.7.
M
ð The automatic rinsing process starts up.
T

10.4 Reprocessing the suction system including suction hoses

Note
Clean the suction hoses after each treatment and disinfect them with DEKASEP‐
TOL Gel daily.