2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit

2.4.6 Servicing the turbine return air filter

Cleaning the turbine return air filter
Note
The liquid collector must be checked weekly.
▶ If there is any liquid in the filter, drain the liquid and replace the O-rings of the mul‐
tiflex couplings.
▶ If necessary, replace the turbine return air filter (Mat. no. 0.200.3098) and clean
the filter holder.
Note
The turbine return air filter is in the same position in the dentist element variant with
S table.

2.4.7 Reprocessing KaVo ERGOcam One

Note
The camera handle cannot be sterilised or disinfected with heat.
Cleaning
Note
Disinfect the camera before cleaning.
Disinfect after cleaning if necessary.
Exterior cleaning by hand
Damage to camera due to improper cleaning.
To avoid the surface or optical system of the camera being destroyed:
▶ Do not use abrasives or aggressive cleaning substances
▶ Do not immerse camera in water or other fluids
▶ Switch main device switch off.
▶ Remove the camera.
CAUTION
17 / 40
Care instructions Primus 1058 Life