Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
▶ Apply KaVo silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no.
0.411.7510) to the O-rings of the joint between the gripping sleeve and cannula.
Use cotton buds for this purpose.
Sterilising the 3-function and multifunction handpiece
Note
The grip sleeve and cannula can be sterilised.
▶ Hold the 3-way or multifunctional handpiece at the gripping sleeve and take off the
cannula with a slight twisting motion.
▶ Pull the gripping sleeve off the valve body.
Sterilization in a steam sterilizer in compliance with DIN EN 13060
▶ Weld the cannula and grip sleeve in sterile goods packaging.
▶ Sterilise with a triple fractionated vacuum (B class steriliser); hold for 4 minutes;
134
°
C ± 1 °C; overpressure: 2.13 bar.
Damage caused by over-heating.
Damage to the gripping sleeve.
▶ Immediately remove the parts from the steriliser after the sterilisation cycle is
completed.
Storage
▶ Reprocessed products should be stored protected from dust with minimum expo‐
sure to germs in a dry, dark and cool place.

2.4.5 Preparation of the instrument hoses

Cleaning and disinfection of instrument hoses
▶ Clean the hoses and couplings with a cloth and water.
Damage caused by spray disinfection.
▶ Do not subject instrument hoses to spray disinfection.
▶ Subject instrument hoses to wipe disinfection.
CAUTION
CAUTION
16 / 40