Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
▶ Pull the latches evenly outward and carefully remove the handpiece holder.
▶ Clean the handpiece holder under running water.
▶ Wipe-disinfect the instrument holder.
Sterilize the instrument holder
Damage caused by sterilisation.
▶ Do not sterilise the instrument holder.

2.4.3 Reprocessing of the instrument arm S table

Note
The instrument arm S table (Mat. no. 1.007.4756) can be removed to facilitate
cleaning.
Cleaning and disinfection of the instrument arm S table
Damage caused by thermodisinfection.
▶ Do not sterilise the instrument arm S table by thermodisinfection.
Note
Neither the rinsing program nor the intensive germ reduction can be carried out
when the instrument arm is removed.
▶ Switch off the unit at the main switch.
CAUTION
CAUTION
12 / 40