2 Reprocessing methods according to DIN EN ISO 17664 | 2.6 Reprocessing of the patient element

2.5.3 Reprocessing of sieve inserts and suction hoses

Note
Check the sieve inserts daily and after each surgery. The sieve inserts must
be replaced at the latest when the suction performance decreases (Mat. no.
10088636).
Cleaning is not recommended for hygienic reasons.
▶ Pull the connector ⑤ off the suction hose.
▶ Clean the sieve ⑥ under running water.
▶ Remove the suction hoses from the holder.
ð
The suction motor starts automatically when the suction tubes are removed
from the holder.
▶ Pull the sieve inserts out of the housing.
▶ Place a cloth under the sieve seat.
▶ If needed, insert a new sieve insert (Mat. no. 10088636) into the seat.
▶ Wipe the suction tubes with disinfectant.

2.6 Reprocessing of the patient element

2.6.1 Reprocessing of the tumbler

▶ Reprocess as specified by the manufacturer.

2.6.2 Reprocessing of tumbler holder and tumbler filler

Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
▶ Make sure that the unit is turned off before removing the tumbler filler or
the spittoon bowl.
▶ Always check the correct seating of tumbler filler and spittoon bowl before
turning on the unit.