Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit

2.4.4 Setting up the 3-function and multifunction handpiece

Note
Wet the O-rings of the gripping sleeves after sterilisation using silicone grease (Mat.
no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510). After replacing the
cannula, the air channel must be blown out before the start of treatment to ensure
that the air exiting during treatment is dry.
Cleaning the triple-function and multifunctional handpiece
Automatic cleaning
Not applicable.
Damage caused by improper cleaning/disinfection.
▶ Do not wash the gripping sleeve and the cannula in the thermodisinfector.
▶ Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶ Do not place the gripping sleeve and the cannula in a disinfectant or ultrasonic
bath.
Manual cleaning
Note
Combine manual cleaning with disinfection.
Cleaning the interior
▶ Leave the gripping sleeve and cannula on the instrument, and activate the air and
water for 60 seconds.
External cleaning
Danger of injury if the valve body is hot (heating elements, high-pressure lamp).
Risk of burn injury.
▶ Switch off the unit at the main switch.
▶ Let the handpiece cool after long use.
Accessories required:
▪ Tap water 30
C ± 5
o
o
▪ Brush, e.g. medium-hard toothbrush
▶ Disassemble the gripping sleeve and cannula.
▶ Detach the key part from the gripping sleeve.
▶ Clean the key part, gripping sleeve and cannula by brushing them under running
water (at least tap water quality).
▶ Immediately remove contamination from materials used in the dental practice (im‐
pression materials, caustic chemicals).
CAUTION
CAUTION
C
14 / 40