At the site of use

2 Setup methods according to DIN EN ISO 17664

2.1 References

The reprocessing consists of the following basic steps:
▪ Cleaning and disinfection ( manually or automatically)
▪ Sterilisation
Note
Comply with national hygiene requirements, e.g. RKI guidelines.
Risk of injury during cleaning of the treatment unit.
Lack of instructions to the cleaning staff and lack of preparation of the treatment unit
can lead to the cleaning personnel sustaining injuries.
▶ Only trained professionals and instructed cleaning personnel may be present in
the treatment rooms.
▶ Position the chair for cleaning and turn the device off.
Product damage due to improper disinfection.
Malfunctions.
▶ Use the disinfectant as prescribed by the manufacturer.
▶ No spray disinfection, perform wipe disinfection only.
▶ Do not immerse product or parts of the product in liquids.
▶ Mop up any spilled cleanser or disinfectant immediately.
Improper disinfection may lead to health hazards and material damage.
Infection hazard to users and patients.
Damage to the sterile product.
▶ No hot air sterilisation, no chemical cold sterilisation, do not sterilise with ethylene
oxide.
Health hazard and property damage due to non-compliance with servicing schedule.
Infection hazard to users and patients.
Product damage.
▶ Comply with servicing schedule.
Note
The preparation methods for the instruments and motors are found in the separate
instructions for use that accompany the instrument and motor packages.

2.2 General instructions for the reprocessing of KaVo treatment units

Please comply with the general reprocessing instructions at all times. Detailed prod‐
uct-specific reprocessing instructions are provided subsequently.
2 Setup methods according to DIN EN ISO 17664 | 2.1 References
CAUTION
CAUTION
CAUTION
CAUTION
▶ Remove any surface contamination using a
disposable cloth/paper towel.