2 Setup methods according to DIN EN ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system
Step
0.1
Start rinsing programme
Note
The rinsing program can be started immediately or the next morning.
▶ Briefly press the "Intensive germ reduction" button to immediately start the proc‐
ess.
ð
The selected rinse program starts.
or
▶ Turn off the device and do not restart it until the next morning.
ð
The cycle is started automatically. The rinsing time follows rinsing program 02.
Note
If the device is not used the entire day (i.e., the instruments are not mounted), the
rinsing program automatically starts the next morning when the device is turned on.
The rinsing programme can be stopped at any time by pressing the "intensive germ
reduction" key.
If the device is turned off and then on again, the rinsing program also stops.

2.8.2 Intensive germ reduction

Note
Observe the instructions for use for the instruments.
Intensive germ reduction prevents the formation of microorganisms during extented
periods when the equipment is out of use. During intensive germ reduction, the affer‐
ent water ducts are automatically filled with a higher concentration of KaVo OXYGEN‐
AL 6. The intensive germ reduction program lasts ca. 45 minutes. The increased hy‐
drogen peroxide concentration acts in the system for at least 30 minutes.
An additional intensive germ reduction must be carried out during extended down‐
times (> 4 weeks).
The need for intensive disinfection is automatically displayed in the top LED of the
"Intensive germ reduction" button at first start-up and after downtimes (weekend, holi‐
days, vacation, etc.).
If an audible signal is issued every 10 seconds and the bottom LED of the "Intensive
germ reduction" button flashes, the KaVo OXYGENAL 6 must be replenished. If a
beep is issued during the filling process (10 times at a one-second interval), the con‐
tainer is full.
See also:
2 Replenish OXYGENAL 6
Note
The suction device cannot be used during the HYDROclean step (4.6).
Description
Waiting position for setting down the handpieces