Short instructions for use 8 Hygiene
Note
The intensive germ reduction bottle must be newly filled prior to every
intensive germ reduction process.
▶ Fill the intensive germ reduction bottle.
See also:
2 Filling the intensive germ reduction bottle
▶ Press the button during the rinse.
▶ Rinse each dental instrument for 20 seconds (without attachment handpieces or can-
nulas), including the triple-function handpiece.
▶ Fill the tumbler 3x in order to rinse and disinfect the entire water-supplying system.
▶ Allow water and KaVo OXYGENAL 6 to act for at least 30 minutes, and preferably
over the weekend.
▶ After the time of exposure is elapsed, fill the tumbler 3x and rinse each dental instru-
ment (without attachment handpieces or cannulas) for 20 seconds.
See also:
2 Servicing Instructions ESTETICA E30

8.4 Reprocessing of the suction system

Note
Clean the suction hoses after each treatment and disinfect them with
DEKASEPTOL Gel daily.
Application of DEKASEPTOL Gel
Note
Regarding the air jet suction system (Venturi), only one aliquot may be
drawn into each hose.
▶ Aspirate one tumbler of cold water with each suction hose.