2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
Cleaning and disinfecting the surfaces of treatment unit / upholstery
Non-colour-fast textiles.
Non-colour-fast textiles can cause permanent discoloration.
▶ Remove discolourations with water immediately.
Approved disinfectants:
▪ CaviCide / CaviWipes (Metrex)
▪ Microcide AF Liquid (Schülke & Mayr)
▪ FD 322 (Dürr Dental)
▪ Incidin Liquid (Ecolab)
▶ Clean the surfaces and upholstery with a soft cloth and water, and wipe-disinfect
them with an approved surface disinfectant.
Damage caused by spray disinfection.
Disinfectant gets into cracks and loosens the paint.
▶ No spray disinfection.
▶ Only disinfect by wiping.
Cleaning and disinfection of the foot control
Damage caused by inappropriate use of water.
Damage to the foot control.
▶ For cleaning, the foot control must not be immersed in water or placed under run‐
ning water.
▶ When wiping it off with a moist cloth, do not touch the contacts of the charge
socket.
▶ Only wipe off. Do not spray.

2.4 Preparation of the dentist unit

2.4.1 Reprocessing of handle and holder pad/anti-slip mat

Note
The handle of the TM table cannot be detached.
CAUTION
CAUTION
CAUTION