Figure 2-41 Sensor Tbls Screen; Table 2-4 Installed Thermistor Sensor Models - Keysight Technologies N432A User Manual

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

2
Operating Information

Table 2-4 Installed thermistor sensor models

[1] DEFAULT is a sensor calibration table where the reference calibration factor and
There are also 17 sensor calibration tables named CUSTOM_0 through
CUSTOM_16. These tables do not contain any data when the N432A is
shipped from the factory.
Editing the sensor calibration tables require the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and calibration factor pairs
4 Save the table
Procedure
First, select the table you want to edit as follows:
1 Press

Figure 2-41 Sensor Tbls screen

74
Table
0
1
2
calibration factors are 100%. This sensor calibration table can be used during the
performance testing of the N432A.
Tables
>
Sensor model
[1]
DEFAULT
478A
8478B
>
to display the Sensor Tbls screen.
Sensor Cal Tables
N432A User's Guide

Advertisement

Table of Contents
loading

Table of Contents