6 Another message will appear, prompting you to confirm whether the
7 Zeroing will start automatically. The Zeroing pop- up message will be
8 Enable the sensor calibration table by pressing
9 Enter the frequency of the RF signal you are measuring by using the
10 Measure V
11 Turn on or reconnect the RF power to the sensor.
12 Measure V
13 Compute the RF power using the following equation:
N432A User's Guide
sensor has been disconnected from the RF source. Press Confirm once
you have verified that there is no RF source connected.
displayed during the zeroing process.
using the up/down arrow key to highlight the CF Table setting field, and
pressing
to display the tables. Use the up/down arrow key to
select the required table and press
to complete the table selection and return to the channel setup
screen.
up/down arrow key to highlight the Frequency setting field, and pressing
. Use the numeric keypad to enter the required value in the
Frequency pop- up, then press either
automatically select and apply the calibration factor from the selected
table based on the entered frequency. Press
the calibration factor value on the Cal Fac setting field.
and V
RF
V
are equivalent to V
COMP
equivalent to V
COMP0
and V
RF
V
are equivalent to V
COMP
equivalent to V
COMP1
=
P
Table
MHz
on the digital multimeters. These V
COMP
and V
RF0
COMP0
– V
.
RF0
on the digital multimeters. These V
COMP
and V
RF1
COMP1
– V
.
RF1
(
2 V
V
V
COMP1
1
COMP0
-----------------------------------------------------------------------------------------------------
4Rε
Operating Information
>
Offsets
to highlight
On
. Press
or
. The N432A will
GHz
>
Offsets
. Calculate V
, which is
0
. Calculate V
, which is
1
) V
2
2
V
+
V
0
0
1
2
, then
Done
>
to check
and
RF
and
RF
81