Calibration Of The Touch Screen - Keysight Technologies E5061B Installation Manual

Network analyzer
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

Installation

Calibration of the Touch Screen

Calibration of the Touch Screen
When E5061B measurement screen appears, you have to calibrate the touch
screen. Follow the procedure described below to calibrate the touch screen.
Step 1. Press
Step 2. Press Service Menu.
Step 3. Press Test Menu.
Step 4. Press Adjust Touch Screen. The touch screen calibration screen
appears.
Figure 1-18
Touch Panel Calibration Screen
Step 5. Touch the symbol on the upper left of the screen with your finger. The
symbols also appears on the upper right, lower left, and lower right of the
screen. Touch all the symbols in that order with your finger.
Touching all the four locations described above with your finger automatically
concludes the touch screen calibration.
If you do not start to perform the touch screen calibration within several
seconds, the previous measurement screen re-appears automatically.
Keysight E5061B Network Analyzer
.
(Figure
1-18)