Keysight Technologies N432A User Manual page 130

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

3
Maintenance
213
214
220
221
222
224
226
230
108
Init ignored
Indicates that a request for a measurement initiation was ignored as the N432A was
already initiated
Example: INIT:CONT ON
INIT
Trigger deadlock
TRIG:SOUR was set to HOLD or BUS and a READ? or MEASure? was attempted,
expecting TRIG:SOUR to be set to IMMediate
Parameter error; Frequency list must be in ascending order
Indicates that the frequencies entered using the MEMory:TABLe:FREQuency command
are not in the ascending order
Settings conflict
This message occurs under a variety of conflicting conditions. The following list gives a
few examples of where this error may occur:
• If the READ? parameters do not match the current settings
• Trying to clear a sensor calibration table when none is selected
Data out of range
A numeric parameter value is outside the valid range for the command
Example: SENS:FREQ 2 HZ
Illegal parameter value
A discrete parameter was received which was not a valid choice for the command. You
may have used an invalid parameter choice.
Example: TRIG:SOUR EXT
Lists not same length
This occurs when SENSe:CORRection:CSET[1]|CSET2:STATe is set to ON and the
frequency and calibration/offset lists do not correspond in length
Data corrupt or stale
This occurs when a FETC? is attempted and either a reset has been received or the
N432A state has changed such that the current measurement is invalidated (for
example, a change of frequency setting or triggering conditions)
N432A User's Guide