N432A User's Guide
Data questionable;Input Overload
The power input to the channel exceeds the maximum range of the sensor
Data questionable;Input Overload ChA
The power input to Channel A exceeds the maximum range of the sensor
Data questionable;Lower window log error
This indicates that a difference measurement in the lower window has given a negative
result when the units of measurement were logarithmic
Data questionable;Upper window log error
This indicates that a difference measurement in the upper window has given a negative
result when the units of measurement were logarithmic
Data questionable;ZERO ERROR
N432A zeroing failed. The most likely cause is attempting to zero when some power
signal is being applied to the sensor.
Data questionable;ZERO ERROR ChA
N432A zeroing failed on Channel A. The most likely cause is attempting to zero when
some power signal is being applied to the sensor.
Data questionable;Measurement Over Range
The voltage measurement failed as the measured data has exceeded the maximum
range
Data questionable;Measurement Under Range
The voltage measurement failed as the measured data is below the minimum range
Hardware missing
The N432A is unable to execute the command because either no sensor is connected or
it expects a compatible thermistor sensor and one is not connected