Maintenance
When the self- test has completed, press
menu.
If the self- test failed, information about the failure will be displayed on
the screen.
For any currently selected bridge resistance value of either 100 Ω, 200 Ω,
300 Ω, or 400 Ω, the internal self- test resistance value of 200 Ω will be
used to perform the self- test.
When you press
connected, you will be prompted again to disconnect the sensor. If you
cancel the sensor disconnect confirmation with the sensor still connected,
tests for V
below.
Figure 3-2 V
100
Run Self Test
followed by
, V
, and V
will be skipped and displayed as shown
RF
COMP
0
V
, and V
tests skipped
RF,
COMP
0
Done
to return to the Self Test
Confirm
even if the sensor is still
N432A User's Guide