A 2.4.8.2 Trigger Level

Sets the active level logic and the switching threshold for the trigger input.
- HIGH|LOW: active level logic
- LLL|HLL: Switching threshold
Default = HIGH LLL

A 2.4.8.3 Number of Measuring Values Displayed

Determines how many measuring values are output after a trigger event.
- 1...16382: Number of measuring values which are displayed after trigger event
- INFINITE|16383: Start the continuous measuring value output after a trigger event
- 0: Stops the continuous output of measuring values
Default = 1

A 2.4.8.4 Software Trigger Pulse

Generating a software trigger. Is the trigger selection is not SOFTWARE, the error message „E43 triggermode
SOFTWARE disabled" is output.
If the command is resent with active measuring value output, the trigger is stopped and the measuring value
output is finished.
A 2.4.9
Set Ethernet interface.
- DHCP: IP address and gateway are automatically requested by DHCP . System looks for a LinkLocal ad-
- STATIC: Set IP address, net mask and gateway in format xxx.xxx.xxx.xxx
Values stay the same if no IP address, net mask, and/or gateway is typed in.
Default = STATIC 169.254.168.150 255.255.0.0 169.254.1.1
C-Box/2A
TRIGGERLEVEL HIGH|LOW LLL|HLL
ƒ LLL = High level logic ==> LO = 0..0.7 Volt, HI = 8..30 Volt)
ƒ HLL = High level logic ==> LO = 0..3 Volt, HI = 8..30 Volt)
TRIGGERCOUNT 0|1...16382|INFINITE|16383
TRIGGERSW
Ethernet
IPCONFIG DHCP|STATIC [<IPAdresse> [<Netmask> [<Gateway>]]]
dress after appr. 30 minutes if no DHCP server is available.