A 2.4.6
- NONE: Sensors are not synchronized, the C-Box/2A runs with its own clock and takes just available sen-
- INTERNAL: C-Box/2A produces Sync impulse
- EXTERNAL: External Sync impulse is looped through to the sensors
Default = INTERNAL LLL
A 2.4.7
The C-Box/2A restarts.
- ALL: Also restart the sensors.
A 2.4.8

A 2.4.8.1 Trigger Selection

Selection of trigger mode
- NONE: No triggering
- EDGE: Level triggering via TRG-IN (Measuring value output depends on TRIGGERCOUNT)
- PULSE: Gate triggering via TRG-IN (continuous measuring value output while TRG-In is inactive.)
- SOFTWARE: Triggering via the command TRIGGERSW (measuring value output depends on TRIGGER-
Default = NONE
C-Box/2A
Synchronization
SYNC NONE|INTERNAL|EXTERNAL [LLL | HLL]
sor values.
ƒ In the case of external triggering it can still be switched between Low Level Logic (LLL) and High Level
Logic (HLL).
ƒ Low Level Logic (0 ... 0,7 to 2,8 ... 30)
ƒ High Level Logic (0 ... 3 to 8 ... 30)
Booting the Controller
RESET [ALL]
Triggering
TRIGGER NONE|EDGE|PULSE|SOFTWARE
COUNT)
Page 87