16.5.2.

Three-point Calibration

During three-point calibration, the MAP 2000/2100 and Macro 900 Meter must
remain switched on. If the Macro 900 Meter is switched off between points, the
calibration process will be aborted and must be re-started from point 1. To
calibrate the ISE electrode follow these steps:
Point 1.
1. Remove the storage cap from the pH electrode, wash the Probe in
distilled water, dry the probe thoroughly then drop the Probe in to the
warm 10ppm solution.
2. Switch the Macro 900 Meter on and stir the probe until the temperature
and NO3 readings are completely stable. A minimum of five minutes is
recommended.
3. Ensure the temperature of the solution is between 20ºC and 40ºC (68ºF -
104ºF).
4. Press the MENU key then select Calibration. The following screen will be
displayed.
5. Select Full Cal. The screen will change to:
6. Select Aux Electrodes. The screen will change to:
The Nitrate (NO3) electrode should have been assigned to AUX socket 2 when it
was fitted. Move the pointer down to 2:NO3 then press the OK or right arrow
key to select.
7. The screen will change to:
© 2013 Palintest Ltd.
Calibration
MacroCal
DO 100%
Full Cal
Calibration
pH/ORP
DO/EC
Aux Electrodes
SELECT ELECTRODE
1:TURB
l 4:N/A
2:NO3
l 5:N/A
3:N/A
l 6:N/A
CALIBRATE NO3
Pt-1? [01/Jan/12]
Pt-2? [01/Jan/12]
Pt-3? [01/Jan/12]
www.palintest.com
V3
Page 100 of 147

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents