Converting Orp Readings To The Hydrogen Scale - palintest Macro 900 Operation Manual

Water quality system
Table of Contents

Advertisement

Macro 900 Operation Manual
13.10.

Converting ORP Readings to the Hydrogen Scale

Electrochemical
hydrogen scale, the convention for which is that the electrochemical potential of
a hydrogen electrode in contact with hydrogen gas at one atmosphere partial
pressure and a solution containing hydrogen ions at unit activity is zero at all
temperatures.
The ORP reference electrode used in Palintest combination electrodes is a
3MPK1 silver chloride type, and exhibits potentials on the hydrogen scale of:
Thus, to refer an ORP potential value measured with the MAP 2000/2100 to the
hydrogen scale, the appropriate value above should be added to the measured
value.
© 2013 Palintest Ltd.
measurements
Temperature
5°C
10°C
15°C
20°C
25°C
30°C
35°C
40°C
www.palintest.com
are
ultimately
Potential
221 mV
217 mV
214 mV
210 mV
207 mV
203 mV
200 mV
196 mV
referred to
the
Page 43 of 147