7. Calibration/Validation

Your Photometer 7500 Bluetooth is delivered with a
calibration certificate validating the performance of the
instrument as it leaves Palintest.
We recommend annual service and calibration of all
photometric instruments in normal use.
The Photometer 7500 Bluetooth also includes an automatic
routine to validate analytical performance using certified
Palintest Check Standards. Accessed via the Mode menu, the
Check Standard Mode provides a field method of ensuring
your instrument is operating within defined specifications and
also a troubleshooting method for unexpected results.
Every Palintest Check Standards set is supplied with certified
values expressed as %T (Transmission), derived from
traceable reference materials. Acceptable tolerances are
defined on the certificate and are automatically specified
within the Photometer 7500 Bluetooth.

7.1 Check Standard Mode

Access Check Standard Mode from the Mode screen.
Highlight Check Standard Mode and press OK
Two choices are offered:
Enter Check Standard Values
Check Standard Measurement insert the Check Standards in
use the up/down keys to
adjust the displayed values to
match the certificate
the defined order to generate
a validation report
35