5.6 Normalisation Temperature

You can set the meter to normalise its
measurements to a temperature of either
25°C or 20°C. Default value is 25°C.
1 Enter advanced setup as described in Section 5.1.
2 Press s or t until [t.nr°C] appears. Press ENTER.
3 Press s or t to select [25.0°C] or [20.0°C].
4 Press ENTER to select and return to the
[t.nr °C] setup function.
5 Press s or t to move to the next setup function
or press CAL to exit to measurement mode.

5.7 Single-Point Calibration

Select "YES" to apply a single calibration value across all ranges.
Select "no" to allow separate calibrations for each range, or to restrict an
individual calibration so that it is applied to one range only. Default value is "Yes".
1 Enter advanced setup as described
in Section 5.1.
2 Press s or t until [S.P.CA] appears.
Press ENTER.
3 Press s or t to select [Yes] or [no].
4 Press ENTER to select and return
to the [S.P.CA] setup function.
5 Press s or t to move to the next
setup function menu or press CAL
to exit to measurement mode.