  • Page 1 Lumiso Chlorine Made in UK Lumiso Chlorine Dioxide Lumiso Ozone User Manual ZIINST64 V1...
  • Page 3: Table Of Contents

    Contents Contents Getting Started Instrument Layout How to Select and Perform a Test Information Messages Keeping the Optics Clean The Settings Menu Using Check Standards Accessing the Results Log Displaying a QR Code Changing Lumiso settings Setting Time and date Adjusting the Display Backlight Using Sample Labels Procedures for Water Testing...
  • Page 4: Getting Started

    Getting Started Thank you for choosing a Palintest Lumiso Photometer. Please take time to read and follow the advice in this manual. If this instrument is used in ways not specified, the protection it provides and its accuracy may be impaired.
  • Page 5 Waterproof USB port Integral Dust Cover Non-Slip Rubberised Base Serial Number L P 1 2 3 4 5 6 7 8 9 Made in theUK Battery Cover with Captive Screw www.palintest.com Non slip rubber feet Integral USB Dust Cover...
  • Page 6: How To Select And Perform A Test

    How to Select and Perform a Test This is the basic procedure for all tests. For specific test details please refer to the Test Instructions. Lumiso Lumiso The test selection menu A Palintest Product A Palintest Product appears when first starting 12:28:49 up Lumiso. 16-07-2021 Cl₂...
  • Page 7 Blanking enables the instrument to set the correct zero value. This ensures the final result will be accurate, even when testing water that is cloudy or coloured. Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Cl₂ (0–5) Cl₂ (0–250)
  • Page 8 Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 The screen will show that the blank is being measured. Pooltest Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 18-01-2021 This indicates that blanking was successful. Pooltest If after blanking, you want to read a different parameter...
  • Page 9 Prepare a 10mL sample by adding reagents according to the test instructions. Place this sample tube in the cell holder. Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 Press read. 16-07-2021 Depending upon test selected, Lumiso will either go to step 9,...
  • Page 10 For Chlorine Tests Only Lumiso Lumiso A choice of free or total chlorine is offered. See test instructions. A Palintest Product A Palintest Product 12:28:49 16-07-2021 Highlight F Cl Highlight F Cl and press F Cl2 (0–5) read read to read and display to read and display T Cl2 (0–5)
  • Page 11 Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 The screen will show that the sample is being measured. Pooltest Lumiso Lumiso The result will be displayed with the following information A Palintest Product A Palintest Product 12:28:49 16-07-2021...
  • Page 12: Information Messages

    A common cause that can trigger any of these messages to be displayed is contamination of the optical windows or stained glassware. See sections Keeping the Optics Clean and How to Get Accurate Results. If messages continue to be displayed, please contact your local Palintest branch or distributor.
  • Page 13: Keeping The Optics Clean

    Keeping the Optics Clean Contamination in the optical cell may effect the accuracy of readings and can trigger an information message to be displayed. Clean optical windows gently with a soft, non-abrasive cloth dampened with water or anti-static foam cleaner. Do not use solvents. Instrument failure due to contamination is not covered by warranty.
  • Page 14: The Settings Menu

    The Settings Menu A number of additional functions and settings can be accessed from within the Settings Menu. Lumiso Lumiso A Palintest Product A Palintest Product 16-07-2021 Cl2 (0–5) The Settings Menu is Cl2 (0–250) accessed from the opening Test Selection Menu.
  • Page 15 Check Standard Mode: Verify instrument performance using NDF check standards Label Mode: Assign a label to your measurement Log Mode: Access the last 50 measurement results QR Log Mode: Access a QR code that contains the current log. Date & Time Mode: Manually edit the date &...
  • Page 16: Using Check Standards

    Using Check Standards Check Standards can be purchased separately for your Lumiso instrument. These standards contain Neutral Density Filters (NDFs) and can be used to verify that the instrument is within calibration and performing correctly. LMC005D LMC001 LMC001 LMC005D Blank LMC005D LMC001 LMC005D...
  • Page 17 Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Check Standard Mode is accessed from the Settings Menu Scroll to the Check Standard Pooltest icon icon and confirm and confirm The wavelength(s) of light the Lumiso Lumiso instrument uses are shown...
  • Page 18 Insert the “Blank” from the set of standards. Ensure that the wavelength Lumiso Lumiso being checked is aligned A Palintest Product A Palintest Product with the mark on the 12:28:49 Lumiso instrument. 16-07-2021 530nm Then press blank...
  • Page 19 Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 The instrument will measure 16-07-2021 the blank and show when this has been successful. A read icon now appears. Pooltest LMP001 Blank Insert Standard A from the set of standards Lumiso...
  • Page 20 The measurement of %T Lumiso Lumiso (Percentage Transmission) A Palintest Product A Palintest Product is shown on screen. 12:28:49 16-07-2021 This value needs to be 530nm compared with that on the 90.8 certificate for the correct wavelength and standard. Pooltest...
  • Page 21 Please clean the cell holder carefully and remeasure the standards. (See ’Keeping the Optics Clean’) If this does not resolve the issue, please contact your local Palintest branch or distributor.
  • Page 22: Accessing The Results Log

    Lumiso stores the most recent 50 test results along with the test time, date and and sample label. The oldest test result will be automatically overwritten once this capacity is reached. Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 The Results Log is accessed...
  • Page 23: Displaying A Qr Code

    (see below). Displaying a QR Code Lumiso Lumiso It is possible to display A Palintest Product A Palintest Product the entire content of the Results Log as a QR code for scanning by other devices. From the Settings Menu...
  • Page 24: Changing Lumiso Settings

    Changing Lumiso Settings Setting Time and Date Changes to the instrument settings are accessed through the Settings Menu. Lumiso Lumiso A Palintest Product A Palintest Product From the Settings Menu From the Settings Menu 12:28:49 16-07-2021 Scroll to the Time and Date...
  • Page 25: Adjusting The Display Backlight

    Adjusting the Display Backlight Lumiso’s screen brightness is adjustable and there are five levels to choose from. Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 From the Settings Menu From the Settings Menu Select the light bulb icon...
  • Page 26: Using Sample Labels

    Using Sample Labels Lumiso Lumiso Sample readings can be tagged A Palintest Product A Palintest Product to indicate a sample name, 12:28:49 16-07-2021 project or sampling location. In the settings menu In the settings menu In the settings menu In the settings menu...
  • Page 27: Procedures For Water Testing

    10 mL line ensure the level 10 mL line ensure the level 10 mL line ensure the level is as shown is as shown is as shown is as shown Use Palintest Photometer reagents. Rapid dissolving and Comparator tablets are not suitable.
  • Page 28 During sample testing or blanking, remove any attached bubbles by capping the tube and rotating as shown. Ensure tubes are dry on the outside before placing them in the instrument. Ensure that your Lumiso Instrument is clean and dry. Place tubes in the instrument with white diamond aligned to the mark on the instrument.
  • Page 29: Blanking

    A dilution tube is available from Palintest to simplify this. Example for a x2 dilution: Fill with sample to x2 line and top up to 100 mL with deionised water. Mix and use this as the new blank and sample for the test.
  • Page 30: Test Instructions

    Test Instructions Depending upon model, any of the following tests may be present in the menu. 0 – 5 mg/L Tablet Chlorine (0-5) Free & Total 0 – 5 mg/L Liquid Chlorine 0 – 250 mg/L (0-250) Tablet High Range 0 - 10 mg/L Chlorine (0-10) Tablet...
  • Page 31 Free & Total Chlorine (DPD Tablet Method) - Cl (0-5) Colourless to Pink Colour Change: 0 – 5 mg/L Range: Rinse tube with sample leaving a few drops. Add one DPD 1 Tablet. Crush tablet to form a paste.
  • Page 32 Fill with further sample to the 10mL line. Stir. Cap the tube. Remove any bubbles by holding the tube and rotating as shown.
  • Page 33 Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Take the Photometer Reading. F Cl2 (0-5) 1.35 mg/L Result = Free Chlorine Pooltest X X X Keep tube and contents to measure Total Chlorine. Add one DPD 3 tablet.
  • Page 34 (0-5), then press read Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Take the Photometer reading. T Cl2 (0-5) 2.55 mg/L Result = Total Chlorine...
  • Page 35 Free & Total Chlorine (DPD Liquid Method) - Cl (0-5) Colourless to Pink Colour Change: 0 – 5 mg/L Range: Rinse tube with sample leaving a few drops. Add 3 drops of DPD A liquid (DPD No1 Indicator) Add 3 drops of DPD B liquid (DPD No1 Buffer) Swirl thoroughly to mix...
  • Page 36 Fill tube with sample to the 10mL line. Swirl to mix. Cap the tube. Remove any bubbles by holding the tube and rotating as shown.
  • Page 37 Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Take the Photometer reading. T Cl2 (0-5) 2.55 mg/L Result = Total Chlorine Pooltest X X X Keep tube and contents to measure Total Chlorine. Add 3 drops of DPD C Liquid.
  • Page 38 (0-5), then press read Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Take the Photometer reading. T Cl2 (0-5) 2.55 mg/L Result = Total Chlorine...
  • Page 39 Chlorine High Range - Cl (0 - 250) Colourless to Yellow to Brown Colour Change: 0 – 250 mg/L Range: Fill tube with sample to the 10mL line. Add, one Acidifying GP tablet and one Chlorine HR tablet. Crush both tablets and stir to mix.
  • Page 40 Replace cap. Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Take the Photometer reading. Cl2 (0-250) 38.8 mg/L Result = Total Chlorine Pooltest...
  • Page 41 Chlorine Dioxide - ClO (0 - 10) Colourless to Pink Colour Change: 0 – 10 mg/L Range: Gently add 10 mL of sample to a clean tube. Avoid any agitation of the sample. Add one Glycine tablet. Crush very gently and stir.
  • Page 42 Gently pour a few drops of this prepared sample into a clean tube. Add one DPD 1 tablet to those few drops of sample. Crush tablet to form a paste. Gently pour the rest of the prepared sample into this tube..
  • Page 43 Cap the tube. Remove any bubbles by holding the tube and rotating as shown. Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Take the Photometer reading. ClO2 (0-10) 4.05 mg/L Result = Chlorine Dioxide Pooltest...
  • Page 44 Ozone – O (0-3) Colourless to Pink Colour Change: 0 – 3 mg/L Range: Rinse tube with sample leaving a few drops. Add one DPD 4 Tablet. Crush tablet to form a paste. Fill with further sample to the 10mL line.
  • Page 45 Stir. Cap the tube. Remove any bubbles by holding the tube and rotating as shown. Lumiso Lumiso A Palintest Product A Palintest Product 12:28:49 16-07-2021 Take the Photometer Reading. O3 (0-3) 1.05 mg/L Result = Ozone Pooltest If chlorine is present, use a glycine tablet, as...
  • Page 46: Technical Specification

    Technical Specification Single wavelength, direct-reading Instrument colorimeter Optics LED light source optical system with narrow band wavelength filters and photodetectors Wavelengths Automatic wavelength selection of 530 nm Wavelength ± 2 nm Tolerance 5 nm Filter Bandwidth 226 x 138 pixel with adjustable backlight LCD Display Results Log 50 results, with date, time and label...
  • Page 47: Changing The Batteries

    Changing the Batteries Please take care when changing the three AA/LR6 batteries in order to maintain the waterproof rating of the instrument. To change the batteries, first disconnect the USB lead. Then use a suitable screwdriver to loosen the captive screw The watertight seal built into the battery cover means that a firm pulling force needs to be...
  • Page 48 The cover will lift and, if necessary, the captive screw can be further loosened so the cover can be removed completely. The batteries are held firmly in place by the clips to enhance Lumiso’s resilience to physical shock. To remove the batteries, push them towards the negative contact and lift the positive end.
  • Page 49 The level of torque applied to this screw has no affect on the performance of the waterproof seal. Thank you for choosing your Palintest Lumiso Photometer. For any further questions or information on Lumiso consumables and accessories please visit www.palintest.com...
  • Page 50 www.palintest.com...