Apply a small amount of silicone grease (supplied with the MAP 2000/2100) to
the threaded section and the O-ring of the AUX Electrode (see photograph).
ENSURE NO GREASE IS APPLIED TO THE GOLD CONTACTS.
Using a clean cloth or tissue paper, polish the gold contacts ensuring they are
completely clean. Carefully insert the electrode into the AUX socket and tighten
firmly until the O-ring is completely compressed.

23.2. Socket Assignment and Calibration

After installation, it is essential to connect the MAP 2000/2100 to a Macro 900
Meter and assign the new electrode type to the relevant AUX Socket. On the
Macro 900 Meter, press the MENU key, then select Setup & Install followed by
Socket Assignment. When the Socket Assignment option has been selected, the
following screen will be displayed.
Using the up and down arrow keys, select the AUX socket you wish to assign
then move the cursor to the right by pressing the right arrow key. When the
cursor has moved to the right of the AUX socket number, use the up and down
arrow keys to select the appropriate electrode type. The tables below show the
available electrode options and the selection that should be made on this
screen:
MAP 2000/2100 type Optical Probes (AUX1 only)
Electrode Part No.
PT 1551
PT 1556
PT 1552
PT 1553
PT 1554
PT 1555
PT 1556
MAP 2000/2100 type ISE Electrodes (AUX2 only)
Electrode
Part
No.
PT 1541
PT 1543
PT 1545
PT 1542
PT 1544
When the desired electrode type is showing, move the cursor back to the left of
the socket number then press OK to send the selection to the MAP 2000/2100.
The socket assignments are stored in the MAP 2000/2100. If you press the ESC
© 2013 Palintest Ltd.
SOCKET
ASSIGNMENTS
1:EMPTY l 4:N/A
2:EMPTY l 5:N/A
3:N/A
Function
Turbidity
Chlorophyll
Phycocyanin (Blue-Green Algae PC)
Phycoerythrin (Blue-Green Algae PE)
Rhodamine WT Dye
Fluorescein Dye
Refined Oil
Function
Ammonium/Ammonia
Chloride
Fluoride
Nitrate
Calcium