Calibrating The Second Point; Calibrating The Third Point; Errors During Calibration; Ph Electrode Efficiency - palintest Macro 900 Operation Manual

Water quality system
Table of Contents

Advertisement

Macro 900 Operation Manual
record book for the Probe. When the offset voltage details have been noted
down, press OK then ESC repeatedly to return to normal reading mode.
Remove the Probe from the calibration bottle, rinse thoroughly in de-ionised
water, shake off any excess and dry the outer sleeve with a soft cloth.
13.5.2.

Calibrating the Second Point

The pH electrode can now be calibrated at either pH 4.01 or pH 10.00. If you
intend to calibrate at both pH 4.01 and pH 10.00, both points must be
calibrated in the same session, i.e. without turning the power off.
If the power is removed after calibrating just one additional point (pH 4.00 for
example), the probe will automatically calculate and save a calibration value for
the uncalibrated third point in order to maintain the electrode's linearity.
To calibrate the second point, fill a calibration bottle with fresh pH 4.01 or pH
10.00 solution and drop the Probe in all the way. Follow the procedure detailed
above, but at step 6, select either pH4.01 or pH10.0, dependent upon the
solution you are using. Wait while the Meter stabilises and calibrates. When the
'Calibrating 100%' screen is displayed, the calibration report will display the
slope for the pH electrode in millivolts (mV) per pH unit. If this slope goes
below 45mV/pH at 25°C, the pH electrode should be serviced. Press OK then
press the ESC key repeatedly to get back to the main display.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water,
shake off any excess and dry the outer sleeve with a soft cloth.
13.5.3.

Calibrating the Third Point

Without switching the Macro 900 Meter off or disconnecting the Probe, fill a
calibration bottle with fresh pH 4.01 or pH 10.00 solution and drop the Probe in
all the way. Follow the procedure detailed above, but at step 6, select either
pH4.01 or pH10.0 dependent upon the solution you are using. Wait while the
Meter stabilises and calibrates. When the 'Calibrating 100%' screen is
displayed, the calibration report will display the slope for the pH electrode in
millivolts (mV) per pH unit. If this slope goes below 45mV/pH at 25°C, the pH
electrode should be serviced. Press OK then press the ESC key repeatedly to
get back to the main display.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water,
shake off any excess and dry the outer sleeve with a soft cloth. Dampen the
sponge in the storage cap with storage solution and fit it to the pH/ORP
electrode. pH calibration is now complete.

13.6. Errors During Calibration

If a problem occurs during calibration, an error message will be displayed.
Refer to
Error! Reference source not found.
handling.

13.7. pH Electrode Efficiency

If the pH electrode becomes worn or clogged, its efficiency and response time
© 2013 Palintest Ltd.
www.palintest.com
in section 10 for error message
V3
Page 39 of 147

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents