26. Appendix 6. ISE Electrodes Detailed Specification

Range
Resolution
Ammonium /
Accuracy
MLD
Ammonia
Interfering
Ions
pH Range
Range
Resolution
Chloride
Accuracy
MLD
Interfering
Ions
pH Range
Range
Resolution
Fluoride
Accuracy
MLD
Interfering
Ions
pH Range
Range
Resolution
Nitrate
Accuracy
MLD
Interfering
Ions
pH Range
Range
Resolution
Calcium
Accuracy
MLD
Interfering
Ions
pH Range
Ammonia readings are calculated from Ammonium, pH and temperature electrode values.
Palintest Ltd. reserves the right to change specifications without notice
Notes:
1. MLD (Minimum Level of Detection) is the minimum value the electrode is
physically capable of measuring.
2. Each ion selective electrode is prone to interference from ions that are
similar in nature to the target ion. The main interfering ions for each
electrode type are listed here. If the water under test contains interfering
ions, the electrode will produce erroneous readings. Ion Selective
Electrodes are not recommended for use in brackish or salt water
due to the high level of interfering ions.
© 2013 Palintest Ltd.
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 – 8,999.9 mg/L
± 10% of reading or 2ppm (whichever is greater)
(1)
(2)
(3)
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 – 19,999.9 mg/L
± 10% of reading or 2ppm (whichever is greater)
(1)
(2)
(3)
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 – 999.9 mg/L
± 10% of reading or 2ppm (whichever is greater)
(1)
(2)
(3)
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 – 29,999.9 mg/L
± 10% of reading or 2ppm (whichever is greater)
(1)
Chloride, Bromide, Fluoride, Sulphate, Chlorate and Perchlorate
(2)
(3)
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0 – 1,999.9 mg/L
± 10% of reading or 2ppm (whichever is greater)
(1)
(2)
(3)
www.palintest.com
0 – 9,000mg/L (ppm)
1.0 ppm
Potassium, Sodium and Magnesium
5 - 8
0 – 20,000mg/L (ppm)
2.0 ppm
Bromide, Iodide, Cyanide and Sulphide
2 - 11
0 – 1,000mg/L (ppm)
0.05 ppm
Hydroxide (OH-)
4 - 8
0 – 30,000mg/L (ppm)
0.5 ppm
3 - 10
0 – 2,000mg/L (ppm)
0.05 ppm
Magnesium, Barium, Lead, Zinc and Sodium
4 - 9
V3