Lens and Sleeve Maintenance

On a daily basis, the lenses on the probe should be wiped over with a soft damp
cloth.
Similarly, the inside of the Probe Sleeve and Sleeve Cap should be kept clean
and free from any deposits that may cause stray fluorescence.
Never use an abrasive cleaner on the inside of the Probe sleeve or cap
as they have been treated with a non-reflective coating which can be
easily damaged. The inside of the sleeve should be wiped over with a soft
damp cloth and non-abrasive detergent.
Always re-calibrate the zero point after cleaning the sleeve or lenses.