Palintest
®
Micro 600
DO Meter
Palintest
®
ZI INST 1240

Advertisement

Table of Contents
loading

Summary of Contents for palintest Micro 600

  • Page 1 Micro600DO_ZI INST 1240 16/9/15 12:57 Page 1 Palintest ® Micro 600 DO Meter Palintest ® ZI INST 1240...
  • Page 2 This manual is written to cover as many anticipated applications of the Micro 600 DO meter as possible. If there are doubts in the use of this meter, please contact us. Our contact details are on the last page of this manual.
  • Page 3: Table Of Contents

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 3 Contents Chapter Page Chapter Page Introduction Probe Maintenance Cap and Electrolyte Replacement Electrolyte Solution Getting Started Description of Keypad Functions Description of LCD Annunciators Troubleshooting Inserting & removing the rubber armour/stand Error Messages Inserting New Batteries Battery Replacement Connecting the Electrode...
  • Page 4: Introduction

    It has a large custom LCD (Liquid Crystal Display) for clear and easy reading. The Micro 600 DO offers measurement of dissolved oxygen (DO) as percentage saturation or concentration. Temperature measurement is also available in degrees Celsius.
  • Page 5: Getting Started

    Other annunciators include “A” (when the ATC function is activated), “CAL” (when meter is in calibration mode) and low battery condition. Active LCD Low Battery Annunciator Temperature Mode display for ATC Annunciator Micro 600 DO milligram per litre (Concentration Mode) parts per million Calibration Percentage (Percentage...
  • Page 6: Inserting & Removing The Rubber Armour/Stand

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 6 Getting Started 2.3 Inserting & removing the rubber armour/stand 1 To remove meter from rubber armour, push out from the bottom edges of meter until it is completely out of boot. Ensure that your electrode cables are not connected.
  • Page 7: Battery Replacement

    3 Insert the mini phono jack of temperature sensor into the socket on the meter as shown opposite. 2.7 Electrode Information The Micro 600 DO includes a galvanic DO electrode Galvanic DO probe PT 148/3 which doesn’t require warm-up. It generates...
  • Page 8: Switching The Meter On

    1 All LCD segments will briefly illuminate. 2 The model Micro 600 DO will display next along with mg/l, ppm, and % annunciators. 3 Next, the meter revision number [r #.##] is shown briefly before beginning measurement.
  • Page 9: Calibration

    See Section 5.2 - Automatic Temperature Compensation See Section 5.9 - Pressure & Salinity Adjustment The Micro 600 DO will accept two % saturation calibration points; 100% using saturated air or air-saturated water, and 0% using zero oxygen solution. When 100% calibration is performed, the corresponding concentration is adjusted simultaneously.
  • Page 10: Temperature Calibration

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 10 Calibration Calibration of the concentration mode will only replace the previous concentration calibration and does not affect the % saturation calibration. To offset your % saturation reading to match another instrument, see Section 5.8 - % Saturation Offset Adjustment. New calibration values will automatically override the existing data.
  • Page 11: Saturation Calibration

    Press CAL as needed to exit calibration without confirmation at any time. 3.3 % Saturation Calibration The Micro 600 DO can be easily calibrated in air. For best DO accuracy, ensure that the barometric pressure value is accurate.
  • Page 12: 0% Calibration

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 12 Calibration 3.3.2 0% Calibration 1 Rinse the probe well with DI water or rinse solution. 2 Press MODE to select % saturation. 3 Dip the probe into zero oxygen solution. Wait for the reading to stabilise. 4 Press CAL.
  • Page 13: Mg/L Or Ppm Calibration

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 13 Calibration 3.4 mg/L or ppm Calibration Performing a 100% saturation calibration will simultaneously calibrate the corresponding mg/L (ppm) concentration value. Therefore, additional mg/L (ppm) calibration isn’t required in most circumstances. If desired, you can perform a calibration adjustment in mg/L (ppm) without affecting your % saturation calibration value.
  • Page 14: Measurement

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 14 Measurement 4.0 Measurement The Micro 600 DO offers Automatic automatic or manual Temperature Compensation temperature compensation. activated See Section 5.2 - Automatic Temperature Compensation. 4.1 Temperature Compensation Insert the phono plug of the probe for Automatic Temperature Compensation (ATC).
  • Page 15: Taking Measurements

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 15 Measurement 4.2 Taking Measurements Follow these general rules when taking measurements: keep the membrane free from contact with solid objects, provide stirring of your solution - this helps to overcome the oxygen consumption of the probe and prevents air bubble entrapment, do not strike against hard surfaces, and do not submerge the cable for extended periods.
  • Page 16: Holding A Reading

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 16 Measurement 4.4 Holding a Reading To hold the displayed reading momentarily, press HOLD/ENTER during Annunciator Blinks when any measurement. The %, meter is in mg/L, ppm or T annunciator “HOLD” mode will blink, indicating that the value is held. Press HOLD/ENTER again to deactivate the HOLD function.
  • Page 17: Setup

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 17 Setup 5.0 Setup There are two setup menus. One menu is derived from the % saturation and temperature modes, the other menu is derived from the mg/L (or ppm) concentration mode. 1 Press MODE to display % or temperature. Press ON/OFF to power off. OR, Press MODE to display mg/L (or ppm).
  • Page 18 Micro600DO_ZI INST 1240 16/9/15 12:57 Page 18 Setup Setup Menu Configuration Menu A.ATC Automatic Temperature Compensation b.DO Select mg/L or ppm units for concentration mode (available from concentration setup only) View the latest calibration data according to the respective setup menu. View the electrode properties FACT View the slope factor OFS View the % saturation offset adjustment.
  • Page 19 Micro600DO_ZI INST 1240 16/9/15 12:57 Page 19 Setup Switch off from % Saturation or Temperature measurement mode. Press and hold MODE key and then switch on. % or Temperature Setup Menu...
  • Page 20 Micro600DO_ZI INST 1240 16/9/15 12:57 Page 20 Setup Switch off from Concentration measurement mode. Press and hold MODE key and then switch on. mg/L (ppm) Concentration Setup Menu...
  • Page 21: Configuration Menu (Cof.1)

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 21 Setup 5.1 Configuration Menu (COF.1) This menu allows the selection/deselection of Automatic Temperature Compensation and selecting the unit of measure (mg/L or ppm) for concentration mode. See sections 5.2 and 5.3 below for further details. 5.2 Automatic Temperature Compensation (A.ATC) Use this menu to change automatic...
  • Page 22: Calibration Data (Cal.2)

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 22 Setup 5.4 Calibration Data (CAL.2) To view the % saturation or mg/L (ppm) concentration calibration data, proceed from the corresponding setup menu. 1 From the setup menu, press ▲▼ to select “CAL.2” for the most recent calibration data. 2 Press HOLD/ENTER to view calibration data.
  • Page 23 Micro600DO_ZI INST 1240 16/9/15 12:57 Page 23 Setup 5 Press HOLD/ENTER again. “LO.mV” is momentarily shown before displaying the electrodes mV output at 0%. Press CAL to exit the setup mode and return to measurement mode at any time.
  • Page 24: Automatic Shut Off (Ato.4)

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 24 Setup 5.6 Automatic Shut Off (AtO.4) Use this feature to conserve batteries. When active, the meter will automatically shut off 20 minutes after the last key press. 1 From the setup menu, press ▲▼ to select “AtO.4”.
  • Page 25: Saturation Offset Adjustment (Ofs.6)

    The probe of the reference meter should be immersed in the same sample at the same depth. 3 Switch off the Micro 600 DO and enter the % saturation setup menu. 4 Press ▲▼ to select “OFS.6”. 5 Press HOLD/ENTER to enter the menu. The...
  • Page 26: Pressure & Salinity Adjustment

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 26 Setup 5.9 Pressure & Salinity Adjustment Use this menu to set barometric pressure and salinity values of the sample to be measured. Use mmHg (A.HG) or kilopascal (A.PA) barometric pressure units, and ppt salinity units (bSAL) from concentration setup menu. For best accuracy, enter the actual salinity value if your samples are at least 0.1 ppt (100 ppm).
  • Page 27 Micro600DO_ZI INST 1240 16/9/15 12:57 Page 27 Setup Pressure & Salinity Adjustment Sequences...
  • Page 28: Probe Maintenance

    Probe Maintenance 6.0 Probe Maintenance The Micro 600 DO probe is a galvanic measuring element which produces an output proportional to the oxygen present in the medium in which it is placed. The galvanic probe design lets you take measurements immediately - without the typical 15 minute wait of polargraphic dissolved oxygen probes.
  • Page 29: Electrolyte Solution

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 29 Probe Maintenance For part numbers, see Section 11 - REPLACEMENTS AND ACCESSORIES. 1 Unscrew the cap counter clockwise from the probe sensing tip. 2 Rinse the probe under running water. 3 Mount the nozzle tip onto the syringe provided. Fill the syringe with the refill solution through the tip of the plastic bottle.
  • Page 30: Troubleshooting

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 30 Troubleshooting Problem Cause Solution a) Batteries not in place a) Ensure batteries are in place No display and making good contact b) Batteries not in correct when b) Re-insert batteries with correct polarity polarity (+/–...
  • Page 31: Error Messages

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 31 Error Messages LCD Display Indicates Cause Solution Need new batteries Clean battery contacts. low battery or battery connection Replace batteries, is bad. noting polarity. Check the value of the Calibration is attempted calibration solution. If when the factory “Err 1”...
  • Page 32: Factory Default Settings

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 32 Factory Default Settings Menu Description Options Default COF.1 Configuration Setup Menu A.ATC Automatic Temperature Compensation YES / NO b.DO DO Concentration Units mg/L or ppm mg/L CAL.2 Calibration Data View only _ _ _ _ ELE.3 Electrode Data Menu FACT...
  • Page 33: Specifications

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 33 Specifications Range 0.00 - 200.0 % % Saturation Resolution 0.1 % Mode Relative accuracy ± 1.5% of Full Scale mg/L (ppm) Range 0.00 - 20.00mg/L or ppm Concentration Resolution 0.01mg/L; 0.01 ppm Mode Relative accuracy ±...
  • Page 34: Replacements And Accessories

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 34 Replacements and Accessories Description Product Code Micro 600 Dissolved Oxygen (DO) Kit PT 1240 Replacement DO Probe (standard cable) PT 148/3 DO membrane maintenance set PT 148/2 Zero Oxygen Calibration Solution, 500ml PT 125/3...
  • Page 35: Dissolved Oxygen Theory

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 35 Dissolved Oxygen Theory 12.0 Dissolved Oxygen Theory Dissolved Oxygen (DO) refers to the volume of oxygen that is contained in water. There are two main sources of DO in water; atmosphere and photosynthesis. Waves and tumbling water mix air into the water where oxygen readily dissolves until saturation occurs.
  • Page 36: Measurement Units

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 36 Dissolved Oxygen Theory Measurement Units One measure of DO in water is parts per million (ppm) which is the number of oxygen molecules (O2) per million total molecules in a sample. Calculating the % Saturation is another way to analyse DO levels.
  • Page 37: Applications

    Micro600DO_ZI INST 1240 16/9/15 12:57 Page 37 Dissolved Oxygen Theory partial pressure of oxygen in the water will be the same as it is in air. Therefore, all you need to know is the temperature of the air in which the probe is placed. By consulting solubility tables for oxygen at the particular barometric pressure and salinity of the water being measured, the corresponding concentration (mg/L or ppm) can be found for air saturated water at the air calibration...
  • Page 38: Warranty

    Warranty 13.0 Warranty The Palintest Micro 600 DO Meter is guaranteed for a period of three years from date of purchase - its associated DO probe is guaranteed for a period of six months from date of purchase. This guarantee excludes accidental damage, or damage caused by unauthorised repair or misuse.
  • Page 39 Micro600DO_ZI INST 1240 16/9/15 12:57 Page 39...
  • Page 40 Micro600DO_ZI INST 1240 17/9/15 08:54 Page 40 For more information on Palintest products, contact your nearest Palintest office or visit our website Palintest Ltd Palintest House, Kingsway, Team Valley, Gateshead, NE11 0NS Tel: +44 (0)191 491 0808 Fax: +44 (0)191 482 5372 sales@palintest.com (UK)