A confirmation screen will be displayed.
If you are sure, press the OK key. If you want to change your mind, press the
ESC key. If you press OK, you will see a message that says CAL RESTORED.
Once factory calibration defaults have been restored, you must carry out a full
calibration of the electrode in question.

10.6. Calibration Data Storage

The MAP 2000/2100 contains its own microprocessor and memory. All
calibration data, including the GLP data, is stored within the Probe's memory.
When a Probe is connected to a Meter, this data is transferred for display and
logging.
This is a major advantage and allows you to use a variety of different Probes
with a single Meter, without the need for re-calibration.

10.7. Calibration Reports

At the conclusion of each successful individual electrode calibration, a single
line Calibration Report is displayed. This report contains the raw output of the
electrode under calibration, uncorrected for temperature.
These values can be recorded and used to track the performance and ageing of
the individual electrodes. Please note however, in order to maximise the value
of this feature, all calibrations must be performed at the same temperature
otherwise the recorded values will not be comparable over time.
No calibration report is generated when using MacroCal.