Entry of New Calibration Code

1
On
start
calibration
button.
Slide the front catch to the right,
and open the instrument case
fully.
2
Press the blue lever, and
insert the contacts of the
calibration chip fully into the
slot revealed.
blue lever.
3
Close the instrument lid and
view the display. Check the new
code number displayed on
screen matches the number on
the sensor packaging. A test
can now be performed.
4
If the calibration procedure was
unsuccessful, an error message
will be displayed.
Do not remove the calibration
chip. Just open and close the lid
again to repeat the calibration.
5
Remove the calibration chip.
up,
change
using
the
Release the
the