FujiFilm DRYPIX 2000 Service Manual page 1043

Medical dry imager
Table of Contents

Advertisement

 8. CHECKING OPERATIONS
8.1
Checking Automatic Density Correction and Conveyance Operations
• Check for film jamming
8.2
Checking Density
• Record density value displayed:
Check that the measured values are within the specified values.
Step
Dmax 3.0
1
fog
2
Density should be between densities of step 1 and step 3.
3
Density should be between densities of step 2 and step 4.
4
0.30 ±0.07
5
0.48 ±0.07
6
0.66 ±0.07
7
0.84 ±0.07
8
1.02 ±0.07
9
1.20 ±0.07
10
1.38 ±0.07
11
1.56 ±0.07
12
1.74 ±0.07
13
1.92 ±0.07
14
2.10 ±0.09
15
2.40 ±0.11
16
2.70 ±0.11
17
3.00 ±0.11
016-201-06E
01.27.2012 FM5920
 Pass
 Fail