Measuring With The Measuring Menus - Ahlborn ALMEMO 5690-2M Operating Instructions Manual

Data acquisition system
Hide thumbs Also See for ALMEMO 5690-2M:
Table of Contents

Advertisement

11. Measuring with the measuring menus

11. MEASURING WITH THE MEASURING MENUS
When the device is switched on for the first
time it displays the menu Measuring points
list (see 11.5.3). This provides a clear over-
view of the whole measuring system. Here you
can check whether the date and time-of-day
are correctly set. If they are not you can enter
the correct values now; (see 10.4 and 10.5).
You are also shown a continuous display of
the measured values for all connected sensors
and measuring channels. You can also, by
means of cursor keys  or  , even assign
other additional functions such as designation,
range, maximum value, and limit values. If you program the cycle timer (see
11.3.2), you can, by pressing <START> , start the first measuring operation
and record measured values cyclically. If a printer or a terminal is connected all
values can also be output online. Having selected the channels you can then
program the measuring points. To select other measured value menus press
.
<ESC>
Menu selection
To ensure that measured values and the asso-
ciated functional values are acquired and dis-
played in your application in the clearest pos-
sible way the 5690-2M system incorporates a
series of preconfigured measuring menus.
These can be selected from the range of
measuring menus
in the number of measuring points (1 to 20), in
the font size used to display measured values
(4, 8, 12 mm), in the choice of bar chart or line
graph, and in the grouping of functions. If
these preconfigured menus do not completely
meet your requirements, you can assemble
your own 3 user menus (U1, U2, U3) from a
range of over 50 functions; (see 11.7).
To activate menu selection press
To select a menu press
To activate the selected menu press
The most important functions for controlling the measuring sequence are al-
ready included in and can also be directly programmed in the measuring
menus. The system also provides special
of programming the sensors and the device and
lar functions.
These can be selected by pressing the keys
30
ALMEMO
; they differ from one another
®
5690-2M
C © REC COM l© ©l R01 * ´´´´´´µµµµ¶
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Meas.points list: Comment
Time: 12:34:56 Date:01.01.04
Cycle-timer:
00: 23.12 °C
01: 11.37 mls Velocity
02: 123.4 mV
10:
53.6 %H
20:
15.2 °C
30:
11.2 glk a.Humidity
START MANU F PRINT ESC
*
ALMEMO 5690-2
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
MEASURING-Menus:
Standard display
U1 Meas. value correction
U2 Averaging
U3 Volume flow
Data logger
Multi channel display
*List of measuring points
Bar charts
Line diagram
_____________________
Menu1 PROGRAMMING-Menus
Menu2 ASSISTANT-Menus
POFF *ON F MENU1 MENU2
<ESC>
or
or
PROGRAMMING menus
ASSISTANT Menus
<MENU1>
00:00:30 nS
Temperatur
U2.4
r.Humidity
Dew point
*
©
...
PROG
for the purposes
for particu-
or
.
<MENU2>

Advertisement

Table of Contents
loading

Table of Contents