Averaging Over Measuring Points - Ahlborn ALMEMO 5690-2M Operating Instructions Manual

Data acquisition system
Hide thumbs Also See for ALMEMO 5690-2M:
Table of Contents

Advertisement

11.4.8 Averaging over measuring points

In all measuring point scans the average value
can also be determined over a number of as-
sociated measuring points. However, for this
average value a function channel with the
measuring range M(n) must be available (see
12.3.9). If you do not wish to program refer-
ence channels and the measuring points to be
averaged begin with M0, you need simply to
program the function channel M(n) to the 2nd
channel of the last connector (e.g. M13) (see
12.3.10). This will then refer automatically to
the series from reference channel 2 (M0)
through to reference channel 1 (M3 = 1st
channel). Other ranges of measuring points can be activated by programming
the reference channels accordingly (see 12.4.6). The function channel can be
configured quickly and easily by means of the wizard menu Averaging .
Sensor channels
4. Chan.
30
31
32
20
21
22
3. Chan.
2. Chan.
10
11
12
1. Chan.
00
01
02
M0 M1 M2 M3
If the sensors are to remain unaffected, the function channel can be pro-
grammed to the device-internal channels (e.g. M19) (see 12.3.10). The default
reference channels are M0 to M1.
Device internal
channels
33
34
38
39
23
24
28
29
14
18
Avg. value M(n)
13
19
03
04
08
09
Difference M01-M00
M4...
M8
AVERAGING
over range of meas. points:
From meas. channel :
00: 234.5 °C NiCr
to meas. channel :
03: 189.7 °C NiCr
Program function channel
to channel:
13: 213.7 °C M(n)
Range:
START MANU M
n=Bk1
M=
M
i
i=Bk2
Example:
n=M3
M13=
i=M0
M
M13=
from M0 to M3
ALMEMO
®
5690-2M
Averaging
M(n)
ESC
/N
M
/N
i
45

Advertisement

Table of Contents
loading

Table of Contents