12. Programming using the programming menus
Device locking
No password, locking with new password
Locking protected by password, enter password
Select locking level for menu and function
Menu Locking the menus
0
None
1
Calibration menu, except password
2
+ programming menus, except recording to memory and output from memory
3
+ recording to memory and output from memory
4
+ wizard menus
5
+ measuring menus, except user menu U1
Fct Locking the functions
0
None
1
Data input, switching on and off
2
+ clear measured data
3
+ start / stop / output measuring operation
4
+ function selection, measuring point selec-
tion
Calibration data management
The device type (with version and serial number) and the sensors (with order
number and serial number) are displayed. With option KL you can enter the
date of the next calibration and the calibration interval in months. If "Calibration
message" is activated then, as soon as the next calibration is due, a message
to this effect will appear when the device is switched on.
82
ALMEMO
®
5690-2M
Password:
Password:
Locking:
and keys
PROG, ON, OFF, ZERO, ADJ
CMEM, CLR, CLRA
START/STOP, MANU, ARRAY, PRINT
PROG, F«, M«
- - - -
****
Menu: 0 Fct: 0