10.4.2 Flow meter adjustment

Open "Flow Meter Adjustment"
"Read from Meter":
"Write to Meter":
After the adjustment the flow sensor is ready for renewed test.

10.4.3 Programming of standard flow curve

Open "Program Flow Meter"
The 59xxxxx number appears from the sensor's type
label.
"Write to Meter":
The flow sensor is now ready for test.
*) METERTOOL automatically configures ULTRAFLOW
page 70). If an alternative meter factor is required, please refer to paragraph 10.4.6 Pulse Configuration DN150-
DN250.
66
Reads data from the connected flow sensor.
Flow curve number - e.g. 5925346 - and meter dimensions appear from the heading.
This number will also appear from the meter's label.
The field "Flow Curve" shows the values of the sensor in question compared to the
standard curve. These values are also shown in the form of a graph.
Writes the correction to the connected flow sensor. The required correction of q
and q
can be entered into the field "Flow Curve Correction".
p
Programs the flow sensor
with the selected standard
*)
flow curve.
5512-385 GB/02.2014/Rev. H1
®