TECHNICAL DESCRIPTION

9.7 Suggested test points

Nom. flow Meter factor
q
p
[m³/h]
[imp/l]
0.6
1.5
2.5
3.5
6
10
15
25
40
60
100
Table 26. Table for ULTRAFLOW
The suggested test parameters are based on EN 1434-5 and q
The test set-ups have been selected on the basis of the following requirements:
Minimum test duration of 3 minutes
Water volumes of q
Water volume of 0.1xq
Water volume of q
These suggested test points can be optimized for each test rig as well as for the test purpose.
Test point
q
q
p
i
[m³/h] [m³/h] [m³/h] [min]
300
0.6
0.006
100
1.5
0.015
60
2.5
0.025
50
3.5
0.035
25
6
0.06
15
10
0.1
10
15
0.15
6
25
0.25
5
40
0.4
2.5
60
0.6
1.5
100
1
including suggested test points, test durations and test quantities.
and 0.1xq
of minimum 10 % of the water volume per hour
i
p
corresponding to minimum 1000 pulses
p
corresponding to minimum 500 pulses
i
5512-385 GB/02.2014/Rev. H1
Test duration
0.1xq
q
q
p
p
i
[min]
0.06
3
20
0.15
3
20
0.25
3
20.2
0.35
3
17.1
0.6
3
20
1
3
20.4
1.5
3
20
2.5
3
20.2
4
3
15
6
3
20
10
3
20
:q
1:100.
i
p
®
ULTRAFLOW
54 and 34
Test quantities
0.1xq
q
q
p
p
i
[min]
[kg]
[kg]
6
30
2
4
75
5
4.8
125
8,4
6
175
10
4
300
20
6
500
34
6
750
50
6
1250
84
6
2000
100
6
3000
200
6
5000
333
0.1xq
p
[kg]
6
10
20
35
40
100
150
250
400
600
1000
51