Start-Up; Measuring Flow; Evacuation - Kamstrup ULTRAFLOW 54 Technical Description

Hide thumbs Also See for ULTRAFLOW 54:
Table of Contents

Advertisement

TECHNICAL DESCRIPTION
Output when using Pulse Transmitter/Pulse Divider with galvanically separated output module (Y=3)
Type
Open collector. Can be connected as three-wire via the built-in pull-up resistance of 39.2 kΩ.
Diagram 17. Block diagram for galvanically separated output module (Y=3).

9.4 Start-up

16 seconds must elapse from start-up to calibration in order to allow a true reading to be reached.

9.5 Measuring flow

To obtain correct flow measurement, the duration of calibration must be minimum 2 minutes.

9.6 Evacuation

®
ULTRAFLOW
must NOT be evacuated (subjected to vacuum).
50
Module Y=3
OC and OD
Max input voltage 6 V
Max input current
0,1 mA
ON condition
U ≤ 0,3 V @ 0,1 mA
OFF condition
R ≥ 6 M
Table 25
5512-385 GB/02.2014/Rev. H1
®