10.4 Application

Flow sensor adjustment.
Before adjusting a sensor you must make sure that the sensor operates satisfactorily in the flow rig in question.
®
See paragraph 9 Calibrating ULTRAFLOW
.
If it is necessary to adjust the sensor more than a few per cent, the sensor is probably defective, or has a different
error, and should not be adjusted.

10.4.1 COM-port selection

Open "Setup"
®
Select a COM-port for ULTRAFLOW
X4.
The USB driver must be installed before connecting
the interface.
The related COM-port will not appear from the list
until the USB interface has been connected.
Select COM-port for Pulse Divider.
Activate "Save" in order to save the selected ports.