Guidance and manufacturer's declaration

Guidance and manufacturer's declaration
– Electromagnetic emission
The medizinisches Bett is intended for use in the electromagnetic environment specified below.
The customer or the user of the medizinisches Bett should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11 (partly)
RF emissions
CISPR 11 (partly)
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
– Electromagnetic immunity
The medizinisches Bett is intended for use in the electromagnetic environment specified below.
The customer or the user of the medizinisches Bett should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrostatic transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short inter-
ruptions and voltage varia-
tions on power supply input
lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
Note: U
is the a. c. mains voltage prior to application of the test level.
T
Complliance
Electromagnetic environment - guidance
Group 1
The medical used bed uses RF energy only for its internal function. Therefore, its RF emis-
sions are very lowand are not likely to cause any interference in nearby electronic
equipment.
Class B
Class A
The medizinisches Bett is suitable for use in all establishments other than domestic and
those directly connected to the public-voltage power supply network that supplies buildings
used for domestic purpose.
Complies
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % U
T
(>95 % dip in U
) for 0.5 cycle
T
40 % U
T
(60 % dip in U
) for 5 cycles
T
70 % U
T
(30 % dip in U
) for 25 cycles
T
< 5 % U
T
(>95 % dip in U
) for 5 sec
T
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % UT
(>95 % dip in UT ) for 0.5
cycle
40 % UT
(60 % dip in UT ) for 5 cycles
70 % UT
(30 % dip in UT ) for 25 cycles
< 5 % UT
(>95 % dip in UT ) for 5 sec
3 A/m
49
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least
30%.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Mains power quality should be that of a typi-
cal commercial or hospital environment. If the
user of the medizinisches Bett requires conti-
nued operation during power mains interrupti-
ons, it is recommended that the medizini-
sches Bett be powered from an uninter-
ruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.