Error On/Off (Function 87); Profibus Configuration (Function 90) - Siemens OXYMAT 64 Operating Instructions Manual

Gas analyzer for measurements of trace oxygen
Hide thumbs Also See for OXYMAT 64:
Table of Contents

Advertisement

5.4.6.16

Error on/off (function 87)

87 Error on/off
S1
Parameter memory
S2
S3
S4
External fault
Figure 5-67
With this function, you switch the message for maintenance requests and faults (see Tables
"Causes of maintenance requests" and "Causes of fault messages") off individually so that
there is no entry in the logbook, no message in the status line of the measured value display
and no external signal sent.
Fault messages which do not apply to this analyzer are identified by the absence of text
following the fault number.
See also
Removing faults (Page 109)
5.4.6.17

PROFIBUS configuration (function 90)

9O Profibus Config.
Address
TAG : OXYMATSIXTYFOUR
Ident number
Relay by PB
Software version: 2.0.O
Boot software
Figure 5-68
This function can only be called if the device contains additional PROFIBUS electronics
(optional board).
With this function, you can set the PROFIBUS station address. The address can be set from
0 to 126.
Under TAG you can display the tag assigned to a device in the ELAN network (or the 16
characters thereof).
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
O
2
...continue
Switching errors on and off
O
2
:126:
:
1:
:off:
: 0.2.0
Configuring PROFIBUS
Operation
5.4 Operating functions
99