Profibus Configuration (Function 90) - Siemens FIDAMAT 6-G Operating Instructions Manual

Gas analyzer for measuring total hydrocarbon concentrations
Table of Contents

Advertisement

With this function, you switch the message for maintenance requests and faults off
individually so that there is no entry in the logbook, no message in the status line of the
measured value display and no external signal sent.
5.11.19

PROFIBUS configuration (function 90)

9O Profibus config.
Adress
Ident number:
Software release: 1.4.O
Figure 5-68
The function can only be called if the device contains additional PROFIBUS electronics
(optional board).
With this function, you can set the PROFIBUS station address. The address can be set from
0 to 126.
The configuration behavior of the device is configured with the
Parameters 0, 1 and 3 can be selected and have the following meaning:
Parameter
0
1
3
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
THC
:126:
O
Configuring PROFIBUS
Meaning
Only the PROFILE ID number is positively acknowledged
Only the device-specific ID number is positively acknowledged
Only the PROFILE ID number for multivariable devices (complex analyzers) are
positively acknowledged.
Operation
5.11 Configuration
"Ident number"
(softkey 2).
5-61