AMI Codes-II TC
Troubleshooting
Error
E020
Photometer dirty
E021
Absorbance too high
E022
Reagent empty
E023
Cleaning Solution
E024
Input active
E026
IC LM75
E028
Signal output open
E030
EEprom Frontend
E031
Calibration Recout
E032
Wrong Frontend
E033
Alarm high pH
E034
Alarm low pH
70
Description
Corrective action
– check process
– clean the photometer, see
Photometer, p. 55
– Appears if disinf. value is too high
– Check dosing system or process
– refill reagents,
see
Refill or replace Reagents, p. 46
– refill cleaning solution
– See If Fault Yes is programmed in Menu
5.3.4, p. 97
– call service
– check wiring on signal outputs 1 and 2
– call service
– call service
– call service
– check process
– check programmed value, see
p. 92
– check process
– check programmed value, see
5.3.1.4.25, p. 93
Cleaning the
5.3.1.4.1,
A-96.250.631 / 280219
Need help?
Do you have a question about the AMI Codes-II TC and is the answer not in the manual?
Questions and answers