AMI Codes-II TC
Maintenance
Standard pH
Display
instructions
A-96.250.631 / 280219
Maintenance
Calibration
Enter
Simulation
Set Time 01.06.04 16:30:00
Enter
Fill System
Cleaning
Calibration
Total chlorine 2
Process pH
Standard pH
1 Rinse and dry the pH sensor and put it into standard 1
2 Standard 1, current value (Progress is shown).
3 Rinse and dry the pH sensor and put it in standard 2
4 Standard 2, current value (Progress is shown)
5 Rinse and dry the pH sensor and put it into the flow cell
Possible error message see
1 Navigate to menu <Maintenance>/
3.1
<Calibration>.
2 Press [Enter].
3 Remove the pH sensor from the
flow cell.
4 Follow the instructions on the dis-
play.
3.1.3
Calibration Errors, p. 66
53