Stop Of Operation For Maintenance - Swan Analytical Instruments AMI Codes-II TC Operator's Manual

Table of Contents

Advertisement

AMI Codes-II TC
Maintenance
6.2.
A-96.250.631 / 280219

Stop of Operation for Maintenance

1 Put the suction lances into a bucket with clean water.
2 Start <Fill system>.
The reagent tubes are flushed with water.
3 Remove the suction lances from the water.
4 Start <Fill system> again.
The water will be pumped out of the reagent tubes.
5 Stop sample flow.
6 Wait until the flow cell is empty.
7 Shut off power of the instrument.
If pH option is installed:
8 Remove the pH sensor [A] from the flow cell.
9 Fill the protective cap [B] with water.
10 Put the protective cap onto the sensor tip.
A
B
A
pH sensor
B
Protective cap
45

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AMI Codes-II TC and is the answer not in the manual?

Questions and answers

Table of Contents