Swan Analytical Instruments AMI Codes-II TC Operator's Manual page 70

Table of Contents

Advertisement

AMI Codes-II TC
Troubleshooting
Error
E001
Alarm high tc1
(Total chlorine 1)
E002
Alarm low tc1
(Total chlorine 1)
E003
Alarm high tc2
(Total chlorine 2)
E004
Alarm low tc2
(Total chlorine 2)
E005
Alarm high cdc
E006
Alarm low cdc
E007
Sample Temp.high
E008
Sample Temp. low
E009
Sample Flow high
68
Description
Corrective action
– check process
– check programmed value, see
5.3.1.1.1.1, p. 91
– check process
– check programmed value, see
5.3.1.1.1.25, p. 91
– check process
– check programmed value, see
5.3.1.1.2.1, p. 91
– check process
– check programmed value, see
5.3.1.1.2.25, p. 91
– check process
– check programmed value, see
5.3.1.1.3.1, p. 91
– check process
– check programmed value, see
5.3.1.1.3.25, p. 91
– check sample temperature
– check programmed value
92
– check sample temperature
– check programmed value
92
– check Inlet pressure
– readjust sample flow
– check program value
5.3.1.2.1, p.
5.3.1.2.25, p.
5.3.1.3.2, p. 92
A-96.250.631 / 280219