AMI Codes-II TC
Program List and Explanations
5.3.4.2
5.3.4.3
5.3.4.4
5.3.4.5
5.4 Miscellaneous
5.4.1
5.4.2
98
Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Cont.:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Output/Control: (relay or signal output):
Cont.:
Controller continues normally.
Hold:
Controller continues on the last valid value.
Off:
Controller is switched off.
Fault:
No:
No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Yes:
Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is
active.
Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6'000 sec
Language: Set the desired language.
Available settings: German /English/French/Spanish
Set defaults: Reset the instrument to factory default values in three
different ways:
Calibration: Sets calibration values back to default. All other
values are kept in memory.
In parts: Communication parameters are kept in memory. All
other values are set back to default values.
Completely: Sets back all values including communication
parameters.
A-96.250.631 / 280219