Setting Advanced
Parameters
Static Density
1-34
ADVAN
In the
option, you can enter parameters for a variety of
specialized operations. You can set parameters for static density,
program tables for steam or natural gas and for multiple K factors,
and modify parameters for the Skan and Skan/Measure techniques.
While following the programming instructions, refer to Figure A-4 on
page A-4 of Appendix A, Menu Maps.To enter the
[F4]
SETUP
press
at the
window.
Use this option to specify basic parameters for mass flow
measurement.
MASS
1. To enter the
submenu, press
2. At the Static Density? prompt, press
NO
press
, the GC868 returns to the
YES
3. If you press
, select the Density Type to calculate Mass Flow:
[F1]
Rho
=
(Fluid Density) or
4. The next prompt depends upon your selection in step 3.
If you select Rho,
a. The GC868 asks for the type of volumetric flow. Press
AcVOL (Actual Volumetric Flow)
select
(Standard Volumetric Flow).
b. Then enter the Fluid Density of the gas to be measured and
[ENT]
press
. (The meter will only accept values from 0.00001-
3
100 lb/ft
.)
If you select Mw,
Enter the Molecular Weight of the gas to be measured and press
[ENT]
.
The GC868 program returns to the
ADVAN
[F1]
ADVAN
at the
window.
[F1]
NO
[F2]
=
or
=
ADVAN
window.
[F2]
Mw
=
(Molecular Weight).
[F2]
or
to select
ADVAN
window.
Programming Site Data
submenu,
YES
. If you
[F1]
to
StVOL