Advertisement

Setting Up Analog Outputs
(cont.)

Option Card Alarms

Programming Site Data
Table 1-5: Error Response Options
Option
HOLD
Holds the last "good" reading.
LOW
Forces the outputs to the low set point.
HIGH
Forces the outputs to the high set point.
OTHER
Allows user to enter 4-20 mA error level in mA.
[F1]
[F4]
7. Use the
-
keys to set up another output or press
select another slot for setup.
For instructions on setting up additional inputs and/or outputs
proceed to the appropriate section. Otherwise, press
I/O
the main
menu prompt.
1. Use the [←] and [→] and the
slot number.
[F1]-[F3]
2. Use the
keys to set up alarm relays A, B, or C,
respectively.
Note: The set up of alarm A is used as an example. Identical
procedures would be used to set up the other alarms.
[F1] = OFF
3. Press
to disable Alarm A and return to the previous
[F2] = HIGH, [F3] = LOW or [F4] = FAULT
prompt, or press
the type for Alarm A.
[F1] = NO
4. Press
for standard operation or
operation.
5. Do one of the following:
FAULT,
If you selected
HIGH
If you selected
a. Use the [←] and [→] and
Output Measurement parameter, as shown in Figure A-2 on
page A-2.
b. Enter a value for the trigger point of the alarm and press
[F1]
[F3]
6. Press
-
to set up another alarm or press
another slot for set up.
For instructions on setting up additional inputs and/or outputs
proceed to the appropriate section. Otherwise, press
I/O
the main
menu prompt.
Display Response
[F1]-[F4]
keys to select the desired
[F2] = YES
proceed to step 6.
LOW,
or
complete the steps below.
[F1]-[F4]
keys to specify the desired
[EXIT]
April 2008
[EXIT]
to
[EXIT]
to return to
to specify
for Failsafe
[ENT]
.
to select
[EXIT]
to return to
1-19