Setting Up The Totalizer/Frequency Outputs - GE DigitalFlow GC868 Programming Manual

Panametrics gas clamp-on ultrasonic flowmeter, 1-channel
Hide thumbs Also See for DigitalFlow GC868:
Table of Contents

Advertisement

April 2008
Setting Up the Totalizer/
Frequency Outputs
Programming a Frequency
Output
1-20
1. Use the [←] and [→] and the
slot number.
[F1]-[F4]
2. Press
to set up outputs A, B, C or D, respectively.
[F1] = OFF
3. Press
to disable output A and return to the previous
[F2] = FREQ or [F3] = TTLZR
prompt, or press
as a frequency or a totalizer output, respectively.
Note: The setup of output A is used as an example. Identical
procedures would be used to set up the other outputs.
To program a frequency output, go to step 4.
To program a totalizer output, go to step 5.
FREQ
[F2]
4. The
output,
proportional to the output measurement. Complete the steps below
to program the frequency output.
a. Use the [←] and [→] and
Output Measurement parameter.
b. Enter the Base value for the low end of the frequency output
[ENT]
range and press
c. Enter the Full value for the high end of the frequency output
[ENT]
range. Press
d. Enter a value between 1 and 10,000 for the Full Scale
[ENT]
Frequency. Press
[F1]-[F4]
e. Use the
keys to select the desired Error Option, the
way the display responds during an error condition. Available
options are shown in Table 1-6 below. Then skip to step 6 on
the next page.
Table 1-6: Error Response Options for Frequency Output
Option
HOLD
Holds the last "good" reading.
LOW
Forces the outputs to the low setpoint.
HIGH
Forces the outputs to the high setpoint.
OTHER
Allows user to enter Hertz error level.
[F1]-[F4]
keys to select the desired
to designate output A
, produces a frequency pulse that is
[F1]-[F4]
keys to specify the desired
.
.
.
Display Response
Programming Site Data